Imaging Device Specifications

This datasheet describes the Reshape Imaging device, a combined incubator-imaging system used to acquire data for the Reshape Microbiology Platform.

Updated over 3 weeks ago

Mechanical characteristics

Screenshot 2024-08-21 at 13.33.33.png

Reshape Imaging device dimensions, front

ris side.png

Reshape Imaging device dimensions, side

Weight: 25 kg

  • The sample trays which hold plates during imaging are 500x330x6mm in outer dimensions.

  • The tallest object that can be placed with clearance under the door/camera in the Imaging Device are approximately 60mm in height.

    • Please note that the device uses a static focus designed for microtiter plates and petri dishes, and so tall objects might not be in focus

Incubation temperature characteristics

  • Minimum Incubation Temperature range (Absolute) from 10°C - 40°C.

  • Minimum Incubation Temperature (Relative to Ambient): +5°C above the ambient temperature.

  • Incubation Temperature Stability: Maintains within ±1°C of the setpoint

  • Incubation Temperature Homogeneity: Uniformity within ±1°C of the setpoint across the chamber.

Power draw and electrical requirements

  • Compatible with both 110V and 230V outlets

  • Max power draw: 150W

  • Max current draw @ 230V: 0.7A

  • Idle power consumption: <20W

  • In the event of a power outage, your Imaging Device will activate the Auto Resume feature, automatically continuing imaging from the current job time once power is restored — no restart required.

  • Please note that any images scheduled during the power outage will be lost.

Environmental Requirements

  • Operating ambient temperature range from 10°C to 35°C.

  • Relative humidity less than 70% RH (non-condensing).

Imaging and camera

  • Resolution: 12 megapixels

Magnification and resolution per well

  • 30 pixels per mm for 90mm Petri dishes

  • 60 pixels per mm for 96-well-plate wells (equivalent to about 10x magnification)

File size

  • Up to 3 MB per image, typically 900kb

File format output

  • Images: JPEG

  • Videos: MP4

Light types

  • Bottom Light

  • Top Light

Fluorescence capability

  • Please be sure to discuss any fluorescence-based application with an application specialist from Reshape so we can help ensure that your application will have adequate signal and signal-to-noise.

  • Standard filter setup optimized for GFP and RFP-like proteins

Long pass emission filter cut-off wavelengths

  • 530 nm

  • 630 nm

Available excitation spectrums

  • Blue: 467nm +/- 5nm

  • Green: 527nm +/- 5nm

Plate format compatibility

  • Formats

    • 90 mm Petri Dish

    • Omni Tray, and 6, 12, 24, 48, 96 and 384-well SBS formats

  • Number of plates

    • 15x Petri Dishes

    • 10x SBS format

Time to image at full capacity

  • Bottom light

    • Petri dishes: <3 minutes

    • 96-well plates: <15 minutes

  • Top light/fluorescence

    • Petri dishes: <3 minutes

    • 96-well plates: <30 minutes

Connectivity and bandwidth requirements

  • The connection should be able to sustain 3 Mb/s upload and download speed per unit. Measuring using an online speed test service gives the best indication.

For more information on connectivity, please consult the network connectivity guide.