Quick Start Guide

Step-by-step setup, initialization, and run instructions for the High-throughput Imaging Device with a built-in screen.

ℹ️
This guide is for the High-throughput Imaging Device with a built-in screen.

If your device does not have a screen, see this guide.

Device overview


Button instructions

ℹ️ The buttons light up when there are available actions.

Preview

Button function

Actions

Primary
Left Side

When device is in 🟢 Ready state, press to start a run


When device is in 🟢 Ready state, hold 3 seconds to enter Cleaning Mode


When device is 🟣 In progress, press to pause


When device is 🟣 In progress, press and hold for 3 seconds to Cancel a run


When device is in 🟡 Cleaning Mode, press to exit

Input/output rotation
Left & right side

When device is in 🟢 Ready state, press to rotate input and output carousels clockwise or counterclockwise

Emergency stop
Left & right side

Press immediately in the event of entrapment or accident, then seek assistance


🔌 Setup

  1. Connect the supplied power and Ethernet cables to the back of the device.

    • For details on power and network requirements, scan the QR code on the front of this document or visit Network Requirements.

  2. Wait for the device to start up.

    • A loading indicator will appear on the screen while the device is booting (~30 seconds).

    • Once ready, the screen will show 🟠 Ready to initialize.


⚡️ Initialize

  1. Ensure input carousel is empty and there is at least one empty petri magazine in the output carousel.

    • Any plates remaining inside the device will be purged into the output magazine during initialization.

    • Check the screen afterwards to see which column to clear.

  2. Press the ⏯️ primary button to begin initialization.

    • The screen will show 🟠 Initializing while the device calibrates (~3 min).

    • Once complete, the screen will show 🟢 Ready.

    • If the screen shows 🔵 Offline, follow the troubleshooting guidelines here.

ℹ️ About initialization

  • Initialization prepares the device to be used for the first time, or after a restart or power cycle.

  • During initialization, the device rotates the input and output carousels and the internal mechanism to safe positions.


▶️ Start a run

  1. Ensure the output carousel has an empty petri magazine in every column, and the screen shows 🟢 Ready.

  2. Fill the input carousel on the left with plates to be analyzed.

    • Load plates into the petri magazines, then slide each magazine into a column of the input carousel until it clicks into place.

    • ❗ All plates must have lids, and plates must not be loaded upside-down.

    • Press the ⬅️ ➡️ rotation buttons on the left to rotate the input carousel and access more columns.

  3. Press the ⏯️ primary button to start a run.

    • The screen will show In progress while the device is imaging.

    • If plates are detected in the output carousel, the run will be automatically cancelled. The screen will show which magazines need to be cleared.

  4. Once the run is complete, remove the plates from the output carousel's petri magazines.

    • The screen will show 🟢 Ready once complete.

ℹ️ Plate categorization / sorting

  • During the run, plates are imaged and analyzed based on the selected analysis in the Reshape Quality platform.

  • Each plate is sorted into a category, with categories mapped to specific output columns and indicated by a consistent color on both the screen and in the UI.

  • The red category indicates an error (e.g. missing barcode, failed analysis). Other colors correspond to the categories configured in the Reshape Quality platform.


⏸️ Pause or cancel a run

  1. To pause a run, press the ⏯️ primary button once.

    • The screen will show 🟣 Paused while the device is paused.

    • To resume the run, press the ⏯️ primary button once more.

  2. To cancel a run, press and hold the ⏯️ primary button for 3 seconds.

    • The device will stop loading new plates and finish imaging any in-progress plates.

    • The screen will show Cancelling while in-progress plates finish imaging.

    • The device may eject plates back to the input carousel. If this happens, a notification will appear on screen — re-seat the ejected plates in their magazine on the input carousel before starting a new run.


🧹 Initiate Cleaning Mode

  1. Hold the ⏯️ primary button for 3 seconds while the device is in 🟢 Ready state.

    • The screen will show 🟡 Cleaning Mode once active.

  2. Follow the full cleaning protocol found here, including PPE requirements and step-by-step disassembly instructions.

  3. Press the ⏯️ primary button to exit 🟡 Cleaning Mode.

    • The device will re-initialize automatically (~3 min).

    • The screen will show 🟢 Ready when the device is ready for a new run.